1 - 8 HS70A Service Manual
WARNING
■ Electric shocks may result if this system, including all of its externally
mounted recording and monitoring devices, is not properly grounded.
■ Never open the cover of the product. Hazardously high voltage flows
through the product. All internal adjustments and replacements must
be made by a qualified Samsung Medison Customer Service
Department.
■ Always check the product's casing, cables, cords, and plugs for
damage before using the product. Disconnect and do not use the
power source if the face is cracked, chipped, torn, the housing is
damaged, or if the cable is abraded.
■ Always disconnect the system from the wall outlet prior to cleaning it.
■ All patient contact devices, such as probes and ECG leads, must be
removed from the patient prior to the application of a high voltage
defibrillation pulse.
■ The use of flammable anesthetic gas or oxidizing gases (N2O) should
be avoided. Doing so may cause an explosion.
■ Avoid placing the system where it is likely to be difficult to operate, or
disconnect.
■ Do not use HF surgical equipment with the system. Any malfunctions
in the HF surgical equipment may result in burns to the patient.
■ The system must only be connected to a supply mains with protective
earth to avoid risk of electric shock.