Chapter 1 Safety 1 - 9
CAUTION
■ The system has been designed for 100-240VAC; you should select the
input voltage of any connected printer and VCR. Prior to connecting a
peripheral power cord, verify that the voltage indicated on the power
cord matches the voltage rating of the peripheral device.
■ An isolation transformer protects the system from power surges. This
continues to operate when the system is on standby.
■ Do not immerse the cable in liquids. Cables are not waterproof.
■ Make sure that the inside of the system is not exposed to or flooded
with liquids. In such cases, fire, electric shock, injury, or damage to the
product may occur.
■ The auxiliary socket outlets installed on this system are rated 100-
240VAC, with a maximum total load of 150VA. Only use these outlets
for supplying power to equipment that is intended to be part of the
ultrasound system. Do not connect additional multiple-socket outlets or
extension cords to the system.
■ Do not connect any peripheral devices not listed in this manual to the
auxiliary socket outlets of the system. It may cause an electrical hazard.
■ Do not touch SIP/SOP and the patient simultaneously. There is a risk of
electric shock from current leakage.