Care and Maintenance
Operating Instructions Arium
®
Comfort I, Comfort II 57
System cleaning is started under Care r Start system cleaning. It takes
approx. 100 minutes and cannot be cancelled.
t Prepare alkaline cleaning solution. Pour at least 1 liter of water (if possible,
RO water) into a 1.5- to 2-liter container and add “Alkaline Cleaner
Solution 1”.
The following figure illustrates the tube guide for the cleaning process.
1
2
3
8
7
5
6
4
Arium
®
Comfort connections
Arium
®
Bagtank connections
Items (1), (2) and (3) only apply to the Arium
®
Comfort II.
t Connect cleaning tubing in front of softener.
t Remove the tubing from the lower port (gray) of the softener cartridge (1).
t Connect the open end of the tubing with the ¼” – ¼” adapter (2) to the
cleaning tubing (3) (included with the accessories).
t Seal the open connection on the softener cartridge (1) with a dummy
plug.
Small amounts of water may escape when connecting the cleaning tubing
at the front of the softener.
r r System cleaning
Prepare alkaline cleaning
solution according to
operating instructions
Cancel OK
Last chance to cancel
r r System cleaning
Connect cleaning tubing in
front of softener
OK