Care and Maintenance
58 Operating Instructions Arium
®
Comfort I, Comfort II
t Disconnect the “Tank Outlet“ (4) tubing from your Arium
®
Bagtank (5).
The tank valve (6) (ball cock) on the “Tank Outlet” tubing (4) must remain
open during system cleaning.
t Attach the “Tank Outlet“ tubing (4), the rinsing (7) and cleaning (3)
tubing to the container with the cleaning solution (8).
t Start System cleaning.
y Rinse the Arium
®
Comfort system with the alkaline cleaning solution for
30 minutes.
You can cancel the cleaning process via the (L) “
Cancel
” key.
Information and Instructions on Disposal “Alkaline Cleaner Solution 1”
This solution can generally be emptied down the drain after the cleaning
process; however, please check with your local authorities first before doing
so to ensure compliance.
t Attach the “Tank Outlet” tubing and the rinsing and cleaning tubing to
the drain.
r r System cleaning
Disconnect “Tank outlet”
tubing from tank
OK
r r System cleaning
Attach the “Tank outlet” tubing
and the rinsing and cleaning
tubing to the container with
the cleaning solution.
OK
r r System cleaning
Start cleaning
Start
r r System cleaning
Cleaning
30 min
Cancel
r r System cleaning
Direct “Tank outlet” tubing
and rinsing and cleaning
tubing towards the drain
OK