Technical Specifications
Operating Instructions Arium
®
Comfort I, Comfort II 91
Feed Water Requirements
Exclusively potable tap water (feed water must meet drinking water standards of the USA, European Union and

3
): 360 ppm
TOC: < 2000 ppb
Colloid index or SDI: < 5
Free chlorine: < 4 ppm
Turbidity: < 1 NTU
Iron (evaluated as total FE content): < 0.1 ppm
Manganese: < 0.05 ppm
Aluminum: < 0.05 ppm
Feed water pH: 4-10
Max. dissolved CO
2
: 
Temperature range: 2 °C – 30 °C
Pressure
1)
: min. 2.0 bar, max. 6.9 bar
1)
Dynamic pressure/flow pressure 100 L/h