Technical Specifications
92 Operating Instructions Arium
®
Comfort I, Comfort II
Water Quality of the Pre-treatment Level
Typical conductivity | resistivity
1)
:
Retention rate up to 99%
Bacteria
2)
< 0.01 CFU/mL
Particles
2)
RO module specifications Retention rate for monovalent ions: up to 96%
Retention rate for polyvalent ions: up to 96%
Retention rate for microorganisms: up to 96%
Retention rate for particulate contaminations:
(see Technical Specifications)
Product water flow performance
9)
up to 5 L/h, and | or 10 L/h at 25 °C
Typical TOC reduction
7)
95%
Ultrapure Water Quality ASTM Type 1
Typical conductivity | resistivity
1)
:
TOC content
4)
(system with UV lamp)
TOC content
4)
(system without UV lamp)
< 5 ppb
Bacteria
2)
< 0.001 CFU/mL
Particles
2)
Endotoxins
8)
< 0.001 EU/mL
RNase concentration
8)
< 1 pg/mL
DNase concentration
8)
< 5 pg/mL
Ultrapure Water Flow Rates
Flow rate with Bagtank 50 and | or 100
(with built-in pump)
up to 2.0 L/min
Flow rate with Bagtank 20, 50, 100
(without pump)
6)
up to 1.0 L/min
Automatic output volume
5)
Volume Scale interval
100 mL – 2 L
2 L – 20 L
20 L – 60 L
0.1 L
1 L
5 L
1)
Depends on feed water quality and temperature
2)
When using a Sterile Plus sterile filter (Sartopore
®
2 150)
3)
Depends on feed water quality, temperature and the status of the RO modules
4)
Determined with municipal water (Goettingen), TOC < 1000 ppb
5)
Depends on connected Bagtank
6)
Bagtank placed on the same level
7)
Depends on the type of organic contamination in feed water
8)
When using an Arium
®
Cell Plus ultrafilter
9)
Depends on device model