3 Safety
18
Version 09/2015, Rev. 1.3 of 02/03/2017
Translation of the original operating manual, WZG6001-e150401
Pos: 43 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0060 Sic herheitshinweise- --------------- --------------- --------------- --------------- @ 26\mod_14053 19021480_68.do cx @ 192031 @ 2 @ 1
3.6 Safety instructions
Pos: 44 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0060-0010 Elektrische Sicher heit @ 26\ mod_1405319022891 _68.do cx @ 192045 @ 3 @ 1
3.6.1 Electrical safety
To reduce the risk of electrical shock, the centrifuge uses a three-wire electrical
cord and plug to connect the equipment to earth-ground. To preserve this safety
feature:
DANGER
Ensure that the wall socket is properly wired and grounded.
Check that the mains voltage agrees with the nominal voltage listed on the
name plate.
Do not place vessels containing liquid on the centrifuge lid or within the safety
distance of 30 cm around the centrifuge. Spilled liquids may get into the
centrifuge and damage electrical or mechanical components.
Work on the power supply system must only be performed by certified
electricians.
Inspect the electrical equipment of the unit regularly. Defects such as loose or
burnt cables must be eliminated immediately.
Pos: 45 /010 Unive rsalmodule/ Leerzeile @ 0 \mod_1202116244 500_0.do cx @ 114 @ @ 1
Pos: 46 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0060-0020 Mechanische Si cherheit (außer 1 -16/1-16K) @ 26\mod_140531 9024314_68.do cx @ 192059 @ 3 @ 1
3.6.2 Mechanical safety
In order to ensure the safe operation of the centrifuge, observe the following:
WARNING
Do not open the lid when the rotor is in motion!
Do not reach into the rotor chamber when the rotor is in motion!
Do not use the centrifuge if it was installed incorrectly.
Do not use the centrifuge without panels.
Do not use the centrifuge if the rotors and inserts show signs of corrosion or
other defects.
Only use the centrifuge with rotors and accessories that have been approved
by the manufacturer. In case of doubt, contact the manufacturer (see chapter
7.3 - "Service contact").
Do not hold your fingers between the lid and the housing when closing the lid.
Risk of crushing!
Defective lid relieving devices could cause the centrifuge lid to fall (contact
the service department, if necessary). Risk of crushing!
Do not hit or move the centrifuge during its operation.
Do not lean against or rest on the centrifuge during its operation.
Do not spin any substances that could damage the material of the rotors and
buckets of the centrifuge in any way. Highly corrosive substances, for example,
damage the material and affect the mechanical strength of the rotors and
buckets.
Stop the centrifuge immediately in the event of a malfunction. Eliminate the
malfunction (see chapter 7 - "Malfunctions and error correction") or inform
the service department of the manufacturer (see chapter 7.3 - "Service
contact").
Ensure that all repairs are performed only by authorised and specialised
personnel.