Version 09/2015, Rev. 1.3 of 02/03/2017
Translation of the original operating manual, WZG6001-e150401
1 Rotor tie down screw
2 Lid
3 Rotor
Fig. 5: Angle rotor with a hermetically sealed lid
Pos: 109 /010 Unive rsalmodule/ Leerzeile @ 0 \mod_120211624 4500_0.do cx @ 114 @ @ 1
Pos: 110 /100 Sigma/1 00 BA Zentr ifugen Sigma (Sta ndardmodule)/0 60 Betrieb_Spin control L+S/060 -0020-0020-0 030 Einset zen von Zubehör 4plät zig bis 2 -16P bis 3-18KS @ 27\mod_1405319 205710_68.do cx @ 192507 @ 4 @ 1
6.2.2.3 Installation of accessories
Only use inserts that are suitable for the rotor (see chapter 11.1 - "Range of
accessories").
All buckets of the swing-out roto need to be installed when spinning.
Always load the axial symmetrical inserts/buckets of the rotors with the same
accessories and fill to avoid imbalance.
Centrifugation with different tube sizes
Working with different tube sizes is possible. In this case, however, it is very
important that axial symmetrical inserts are identical.
Fig. 6: Permissible and impermissible loading of a swing-out rotor with different tube sizes
(example illustrations)