7 Malfunctions and error correction
Version 09/2015, Rev. 1.3 of 02/03/2017
Translation of the original operating manual, WZG6001-e150401
Pos: 158 /100 Sigma/1 11 BA Zentr ifugen Sartorius (Proje kte)/Centr isart G-16/070 S törungen und Fehlersuche/070 -0010-0010 Notentr iegelung G-1 6 @ 37\mod_1437113064917_6 8.docx @ 2724 67 @ 3 @ 1
7.1.1 Emergency lid release
In the event of a power failure, it is possible to manually open the centrifuge lid.
Switch off the mains power switch and disconnect the power cord from the
socket.
Remove the plug (see figure, item 1) from the opening at the side panel, e.g.
with a screwdriver.
Fig. 25: Position of the opening for the emergency lid release
The stopper is connected with a string. Pull the string in order to release the lid.
Then, reinsert the stopper into the opening.
Do not unlock or open the lid unless the rotor is at a standstill.
If the lid is opened via the emergency lid release system during a centrifuge run,
the centrifuge will be switched off immediately and decelerate in an unbraked
manner.