6 Using the centrifuge
34
Version 09/2015, Rev. 1.3 of 02/03/2017
Translation of the original operating manual, WZG6001-e150401
Short run
A short run can be started if no run is active.
Keep the start key pressed during the short run.
During the short run, the centrifuge accelerates with the acceleration curve 9
(maximum) until the maximum permissible speed of the rotor is reached. The
runtime is counted and in the "Time" field the indication "Short" is displayed.
When the start key is released, the centrifuge decelerates to a standstill based on
the maximum deceleration curve.
When the short run is completed, the original parameters (curves, runtime, and
final speed) are restored and displayed.
Fig. 14:
Pos: 131 /010 Unive rsalmodule/ Leerzeile @ 0 \mod_120211624 4500_0.do cx @ 114 @ @ 1
Pos: 132 /100 Sigma/1 01 BA Zent rifugen Sartorius (Standardmodule) /060 Betrie b/Spincontrol L/060 -0030-0020-00 81 Rotoraus wahl - ohne Te mp Sartorius @ 37\mod_14373744 57289_68.docx @ 272920 @ 4 @ 1
6.3.2.7 Rotor selection
This field shows the rotor that is currently being used.
NOTE
The rotor selection can only be changed when the centrifuge is at a standstill.
Select the "Rotor" field and confirm the selection. The indication "set" flashes in
the activated status.
Select the desired rotor.
Confirm the input. The selected rotor will be adopted.
Fig. 15: Preselection of a rotor or a rotor/bucket combination
Automatic rotor identification system
If the centrifuge is equipped with an automatic rotor identification system, the
input mode will be activated automatically if the system detects a different rotor
than the rotor that is set. If the input mode is aborted, this bucket will be stored
nonetheless.
This prevents the maximum permissible speed from being exceeded.