8 Maintenance and service
46
Version 09/2015, Rev. 1.3 of 02/03/2017
Translation of the original operating manual, WZG6001-e150401
Pos: 170 /100 Sigma/1 00 BA Zentr ifugen Sigma (Sta ndardmodule)/0 80 Wartung und I nstandhaltu ng/080-0010-0 020 Zubehör (a ußer 8K) @ 28\m od_14053206 18247_68.docx @ 194845 @ 3 @ 1
8.1.2 Accessories
CAUTION
For the care of the accessories, special safety measures must be considered as
these are measures that will ensure operational safety at the same time!
Immediately rinse off the rotor, buckets, or accessories under running water if
they have come into contact with any liquids that may cause corrosion. Use a
brush for test tubes in order to clean the bores of angle rotors. Turn the rotor
upside down and allow it to dry completely.
Clean the accessories outside the centrifuge once a week or preferably after
each use. Adapters should be removed, cleaned and dried.
CAUTION
Do not clean the accessories in a dishwasher!
Cleaning in a dishwasher removes the anodised coating; the result is cracking in
areas that are subject to stress.
If the rotors or accessories have been contaminated with toxic, radioactive, or
pathogenic substances, clean them immediately with a suitable
decontamination agent (depending on the type of contamination). Take
suitable precautions for your own safety if there is a risk of toxic, radioactive,
or pathogenic contamination.
Dry the accessories with a soft cloth or in a drying chamber at approx. 50°C.