5 Set-up and connection
Version 09/2015, Rev. 1.3 of 02/03/2017
25
Translation of the original operating manual, WZG6001-e150401
Pos: 89 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/05 0 Aufstellung und Ans chluss/050 Aufstellung und Ans chluss==== ============================== @ 27\mod_14053191730 71_68.do cx @ 192381 @ 1 @ 1
5 Set-up and connection
Pos: 90 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/05 0 Aufstellung und Ans chluss/050 -0010 Aufst ellort (außer 8 KS) @ 27\mod_1 40531917427 6_68.docx @ 19 2395 @ 2 @ 1
5.1 Installation site
Operate the centrifuge only in closed and dry rooms.
All the energy supplied to the centrifuge is converted into heat and emitted to the
ambient air.
Ensure sufficient ventilation.
Keep a safety distance of at least 30 cm around the centrifuge so that the vents
in the centrifuge remain fully effective.
Do not subject the centrifuge to thermal stress, e.g. by positioning it near heat
generators.
Avoid direct sunlight (UV radiation).
The table must be stable and have a solid, even surface.
Attention: During transport from cold to warmer places, condensational water
will collect inside the centrifuge. It is important to allow sufficient time for
drying (min. 24 h) before the centrifuge can be used again.
Pos: 91 /010 Unive rsalmodule/ Lee rzeile @ 0 \mod_1202116244 500_0.do cx @ 114 @ @ 1
Pos: 92 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule )/050 Aufstellung und Anschluss/050 -0020 Energi eversorgung-- --------------- --------------- ----------------- ----- @ 27\mod _1405319175 630_68.docx @ 192409 @ 2 @ 1
5.2 Power supply
Pos: 93 /100 Sigma/10 1 BA Zentrifugen Sartorius (Standardmodule )/050 Aufstellung und Anschluss/0 50-0020-0 011 Anschlussa rt (Kaltgeräte stecker+ Schutzschalter) Sartorius @ 37\mod_14 36955071190_68. docx @ 269181 @ 3 @ 1
5.2.1 Type of connection
DANGER
The operating voltage on the name plate must correspond to the local supply
voltage!
Sartorius centrifuges are units of protection class I. The centrifuges of this model
series have a three-wire power cord with an IEC C13 connector. They are equipped
with a mains power switch with an integrated thermal circuit breaker.
Switch the unit off by actuating the mains power switch.
If it has tripped, let the circuit breaker cool for approximately 2 minutes.
Switch the unit on.
The centrifuge is now ready for operation.
Pos: 94 /010 Unive rsalmodule/ Lee rzeile @ 0 \mod_1202116244 500_0.do cx @ 114 @ @ 1
Pos: 95 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule )/050 Aufstellung und Anschluss/050 -0020-002 0 Sicherungen bauseits (außer 3 -30KS) @ 7\m od_1306312302397_ 68.docx @ 43513 @ 3 @ 1
5.2.2 Customer-provided fuses
Typically, the centrifuge must be protected with 16 Amp L or B fuses that are to be
provided by the customer.
Pos: 96 /010 Unive rsalmodule/Abschnitt swechsel @ 0\mod_120212451 4062_0. docx @ 418 @ @ 1
Pos: 97 /010 Unive rsalmodule/Seite nwechsel @ 0\mo d_120211624 4312_0.docx @ 105 @ @ 1