7 Malfunctions and error correction
Version 09/2015, Rev. 1.3 of 02/03/2017
Translation of the original operating manual, WZG6001-e150401
Pos: 155 /100 Sigma/1 00 BA Zentr ifugen Sigma (Sta ndardmodule)/0 70 Störungen un d Fehlersuche/0 70 Störungen u nd Fehlersuche== =========================== @ 28\mod_1405320576072_68.do cx @ 194663 @ 1 @ 1
7 Malfunctions and error correction
Pos: 156 /100 Sigma/1 00 BA Zentr ifugen Sigma (Sta ndardmodule)/0 70 Störungen un d Fehlersuche/0 70-0015 Allgeme ine Stör ungen_Spincontrol L @ 28\mod _1405320581367_68. docx @ 194733 @ 2 @ 1
7.1 General malfunctions
Error messages are displayed as "Error" followed by a code number. If the acoustic
signal is activated, it sounds when the error message is displayed.
Eliminate the source of the problem (see table below).
Acknowledge the error messages by pressing the lid key.
Error messages can be eliminated by pressing the lid key. The error itself will not
be eliminated, but the centrifuge can be operated again.
No indication on the display
No power in the mains supply
Check fuse in the mains supply
Power cord is not plugged in
Plug in power cord correctly
Reactivate temperature fuse (see
chapter 5.2.1 - "Type of connection")
Switch mains power switch on
Centrifuge cannot be started:
start key LED is not illuminated
Power off/on. If the error occurs
again, contact service
Centrifuge cannot be started:
lid key LED flashes
The lid lock is not closed correctly
Open and close the lid. If the error
occurs again, contact service
Centrifuge decelerates during
operation
Brief mains power failure
Press start key in order to restart the
centrifuge
Power off/on. If the error occurs
again, contact service
Centrifuge decelerates during
operation, imbalance message is
displayed
Improper loading
Centrifuge is inclined
Drive problem
Centrifuge was moved during run
Balance load and restart the
centrifuge. If the error occurs again,
contact service(see chapter 7.1.1 -
"Emergency lid release")
Ungreased load- bearing bolts
Clean and grease load- bearing bolts
Lid lock has not released
Unlock the lid manually and contact
service
Clean the lid seal and apply talcum
powder
Temperature value cannot be reached
(only for refrigerated centrifuges)
Condenser dirty (only air-cooled
units)
Clean the condenser. If the error
occurs again, contact service
Pos: 157 /010 Unive rsalmodule/Seite nwechsel @ 0\m od_12021162 44312_0.docx @ 105 @ @ 1