6.3.3 Program mode
A program contains all the data that are required for a centrifuge run. Certain
sedimentation results can be repeated under identical conditions.
Programs can be saved, loaded, executed, and edited, when the centrifuge is at a
standstill.
A maximum of 50 programs can be stored under the numbers 1 - 50.
"--" means that the values that are currently set are not a stored program.
The programs can be protected against unauthorised use, modification, or deletion
with the aid of an input lock (see chapter 6.3.2.13 - "Input lock").
Pos: 147 /010 Unive rsalmodule/ Leerzeile @ 0 \mod_120211624 4500_0.do cx @ 114 @ @ 1
Pos: 148 /100 Sigma/1 01 BA Zentr ifugen Sartorius (Sta ndardmodule) /060 Betrie b/Spincontrol L/060 -0030-0030-00 11 Progra mm speichern - ohne Temp Sartor ius @ 37\mod_143 7374504107_ 68.docx @ 2731 00 @ 4 @ 1
6.3.3.1 Saving a program
This function is only available if the centrifuge is at a standstill.
Enter the parameters that are to be included in the program.
Select the menu item "save Prog" and confirm the selection. The indication
"save" flashes in the activated status.
Select a random storage location from the program selection list. Free storage
locations are indicated by a flashing display. Any storage locations that are
already occupied will be overwritten during the saving process.
Save the program in the desired location.
The program is now saved.
The functions "Auto Open", "Buzzer", "ProgLock", and "Cycles" cannot be entered
as part of a program.
Fig. 23: Saving a program