EasyManua.ls Logo

Sartorius Centrisart G-16 - Cycle Display (Cycles)

Sartorius Centrisart G-16
72 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6 Using the centrifuge
Version 09/2015, Rev. 1.3 of 02/03/2017
37
Translation of the original operating manual, WZG6001-e150401
Pos: 142 /100 Sigma/1 01 BA Zentr ifugen Sartorius (Sta ndardmodule) /060 Betr ieb/Spincontrol L/060 -0030-0020-01 31 Zyklenanzeige L - ohne Temp Sartorius @ 37\mod_14373710 96266_68. docx @ 272811 @ 4 @ 1
6.3.2.12 Cycle display ("Cycles")
In order to activate the cycle display:
Select the "Cycles" symbol with the cursor and confirm the selection. The
symbol is displayed and "set" flashes in front of the rotor display.
All of the rotors can be selected by turning the function knob. The cycles of the
selected rotor are displayed.
Press the function knob in order quit the cycle display.
Fig. 20: Cycle display for rotor A7B (example)
Reaching the maximum number of cycles
When the maximum number of cycles is reached for a rotor or bucket, a
corresponding warning signal will be displayed every time that the centrifuge is
started: the start key, lid key, and entire display will flash.
Fig. 21: Display when the maximum number of cycles is reached (flashing)
When the start key is pressed, "CYCLES" will be displayed. The centrifuge will not be
started and the display will not return to its normal state until the start key is
pressed again.
WARNING
When the maximum number of cycles of the rotor is reached, the rotor must be
replaced immediately for safety reasons (see chapter 3.6.7 - "Service life of rotors
and accessories").
The cycle display will be reset after the rotor has been replaced by the service
department of Sartorius (see chapter 7.3 - "Service contact").
Pos: 143 /010 Unive rsalmodule/Seite nwechsel @ 0\m od_12021162 44312_0.docx @ 105 @ @ 1