3 Safety
Version 09/2015, Rev. 1.3 of 02/03/2017
21
Translation of the original operating manual, WZG6001-e150401
Pos: 58 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0070 Sic herheitseinrichtungen ------------ --------------- -------------- @ 27\mod_1405319 034809_68. docx @ 192157 @ 2 @ 1
3.7 Safety devices
Pos: 59 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0070-0010 Deckelver riegelung @ 27\mod_1 40531903632 1_68.docx @ 19 2171 @ 3 @ 1
3.7.1 Lid lock device
The centrifuge can only be started when the lid is properly closed. The electrical
lock must be locked. The lid can only be opened when the rotor has stopped. If the
lid is opened by way of the emergency release system during operation, the
centrifuge will immediately switch off and decelerate brakeless. If the lid is open,
the drive is completely separated from the mains power supply, i.e. the centrifuge
cannot be started (see chapter 7.1.1 - "Emergency lid release").
Pos: 60 /010 Unive rsalmodule/ Leerzeile @ 0 \mod_1202116244 500_0.do cx @ 114 @ @ 1
Pos: 61 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0070-0020 Stillstandsüberwa chung @ 27 \mod_140531 9037803_68.do cx @ 192185 @ 3 @ 1
3.7.2 Standstill monitoring system
Opening of the centrifuge lid is only possible if the rotor is at a standstill. This
standstill is checked by the microprocessor.
Pos: 62 /010 Unive rsalmodule/ Leerzeile @ 0 \mod_1202116244 500_0.do cx @ 114 @ @ 1
Pos: 63 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0070-0031 Systemkontr olle Spincontrol L @ 27\mod_14053 19040423_68.do cx @ 192213 @ 3 @ 1
3.7.3 System check
An internal system check monitors the data transfer and sensor signals with regard
to plausibility. The system continuously performs a self-check and identifies
malfunctions. Error messages are displayed as "Error" followed by a code number
(see chapter 7.2 - "Table of error codes").
Pos: 64 /010 Unive rsalmodule/ Leerzeile @ 0 \mod_1202116244 500_0.do cx @ 114 @ @ 1
Pos: 65 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0070-0040 Schutzleiterpr üfung Spincontrol L+S +Easy+U niversal @ 27\m od_1405319043129_ 68.docx @ 192241 @ 3 @ 1
3.7.4 Earth conductor check
For the earth conductor check, there is an equipotential bonding screw on the rear
panel of the centrifuge (see chapter 2.1.1 - "Functional and operating elements").
An earth conductor check can be carried out by authorized and specialized
personnel using a suitable measuring instrument. Please contact the service
department (see chapter 7.3 - "Service contact").
Pos: 66 /010 Unive rsalmodule/ Leerzeile @ 0 \mod_1202116244 500_0.do cx @ 114 @ @ 1
Pos: 67 /100 Sigma/10 0 BA Zentrifugen Sigma (Sta ndardmodule)/03 0 Sicherheit/030- 0070-0051 Unwuchtüber wachung Spincontr ol L+Easy+Univ ersal @ 27\ mod_1405319047031 _68.docx @ 192 283 @ 3 @ 1
3.7.5 Imbalance monitoring system
The indication "Imbalance" in the rotor field and, in some cases, also a sound signal
indicate that the centrifuge is in the impermissible imbalance range.The drive will
be switched off in the acceleration phase or during the run.
Pos: 68 /010 Unive rsalmodule/ Lee rzeile @ 0 \mod_1202116244 500_0.do cx @ 114 @ @ 1
Pos: 69 /100 Sigma/10 1 BA Zentrifugen Sartorius (Sta ndardmodule)/ 030 Sicherheit/030 -0070-007 0 Rotorüberwa chung Sartorius @ 38\mod_14 39880024933 _68.docx @ 276 092 @ 3 @ 1
3.7.6 Rotor monitoring system
When a rotor number number is selected, the computer will automatically check
whether the entered speed or the entered gravitational field are permissible for the
selected rotor.
Pos: 70 /010 Unive rsalmodule/Seite nwechsel @ 0\mo d_120211624 4312_0.docx @ 105 @ @ 1