Procedure
t Remove the cap.
t Slide the menu access switch to the left (= “open" position).
If the device is part of a verified weighing facility, this will only be possible if the
verification seal is broken. The weighing equipment must then be verified again.
t Activate the Service mode, see “Service Mode."
t Select weighing platform “WP 1" in the “Device Parameters" menu item.
t If the “Internal" setting is not already activated (marked by “o"), select the setting
using the “
Q" or “q" soft key and confirm with “O."
y The menu for the “WP-1 INTERNAL" device parameters is displayed.
t Use the “Q" or “q" soft key to select and open the
“
Calibration/adjustment" menu using “O."
t Use the “Q" or “q" soft key to select and use “O" to open the “Geographical
data" menu.
t Use the “O" soft key to confirm “Input parameters."
Entering Geographical Latitude and Elevation
t Use the “Q" or “q" soft key to select the corresponding input field.
t Enter the number via the keypad and confirm using the “l" soft key.
y The next input field is selected.
In this example, the geographical data are entered for the respective platform as a value
pair “Latitude" and “Altitude." After this data was saved and the scale
returned to weighing mode, this pair of values is displayed again the next time the
display menu is opened. The input field for gravitational acceleration is empty.
In this example, the value for gravitational acceleration is entered for the place of
installation. The fields “
Latitude" and “Altitude" are invalid. The set value is
then re-displayed after it is saved and the input menu is re-opened. If you exit the Setup
menu and then open the Service mode, the set value for gravity is no longer displayed.
28