EasyManua.ls Logo

SBM SB-3000 - 8.6.1. CF level

SBM SB-3000
253 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
84
[6] In USB Memory Stick, Log DATA file will be
created as right screen.
8.5.0. Machine setting
8.5.1. Set date/time
8.5.2. Set A/S call: Setting A/S phone number
8.5.3. Ethernet: You can manage LAN here and can set IP address.
8.6.0. Value setting
8.6.1. CF level
In this menu, Service Man can set the levels such as UV (Ultra-Violet) Sensor, MR (Magnetic Reader)
Sensor and IR (Infrared) Sensor required detecting each currency. The setting method is as follows.
[How to set]
[1] Select Currency. ()
[2] Move each circle and set the levels ()
Move the circle of the first bar and set UV level.
Move the circle of the second bar and set IR
level.
Move the circle of the third bar and set MR
level.
(It is possible to set each level in detail by using “ ” key)◀▶
[3] Press [START/STOP] Key to save the level.
The lower number is less sensitive. If the machine detects CF note as real notes, increase level by one to
set the level. If the machine detect real note as CF note, decrease level by one to set the level.
If the level is ‘0’ (Zero), the sensor is deactivated.
(FIG 8-13)
(FIG 8-14)

Table of Contents