98
9.2.0. TAPE Calibration
This Tape Calibration is for Tape Sensors.
You should use the Tape calibration sheet provided from SBM as below
<CAP SHADING SHEET>
(180x 105 mm)
[Procedure]
[1] Touch on TAPE Calibration MENU
[SEVICE MENU _ Machine management/
Calibration/ TAPE calibration]
[2] Open the REAR DOOR B, and Open the REAR MIDDLE COVER
[3] Put the sheet in this machine as right picture
[4] Close the REAR MIDDLE COVER and Close the REAR DOOR B
(FIG 9-06)
(FIG 9-07)
(FIG 9-08)