Average NG: if the average voltage is
more than 1V ± 10%.
1. Need to clean CIS sensors (CIS1 and Opt CIS)
2. Need to check CIS’s defect
3. Need to check Image board’s defect
4
Uniformity NG: If the uniformity is less
than 65%
5
NG if each R.G.B On time is more than
150.
6
NG if all R.G.B On time is 255
▶ Even if the all of pixel is on, target
voltage couldn’t be achieved.
Need to check the CIS receive sensor