100
9.3.0. UV Calibration
This UV Calibration is for UV Sensors.
You should use the UV calibration sheet provided from SBM as below
<UV SHADING SHEET: SHINWOO UV SHADING V1.0>
[Procedure]
[1] Touch on UV Calibration MENU
[SEVICE MENU _ Machine management/
Calibration/ UV calibration]
[2] Open the REAR DOOR B, and Open the REAR MIDDLE COVER
[3] Put the sheet in this machine as right picture
165 mm
75 mm
(FIG 9-17)
(FIG 9-18)