EasyManua.ls Logo

SciCan Statim 5000 - Removing a Detachable Power Cord; Replacing;Refitting a Detachable Power Cord

SciCan Statim 5000
271 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 155
STAT
IM
2000/2000S Service Guide
96-106775 Rev 5.0
7. Electrical and Electronic Components
STATIM 5000/5000S/5000 G4 Service Guide
Removing a Detachable Power Cord
To remove a detachable power cord (1) connected to the unit through a panel mount A.C. inlet
receptacle, proceed as follows (see Figure 20):
1. Turn the power switch (2) OFF, and unplug the power cord from the wall outlet.
2. Disconnect the power cord from the panel mount A.C. inlet receptacle (3).
Replacing/refitting a Detachable Power Cord
Use replacement cords with appropriate safety agency ratings and approvals only.
To replace/reinstall a detachable power cord (1), proceed as follows (see Figure 20):
1. Select the appropriate replacement cord from the spare parts list.
2. The A.C. inlet receptacle (3) is keyed. Note the shape of the cord connector and the
corresponding shape of the receptacle. Plug in the connector.
3. Plug the cord into the wall receptacle and turn the power switch (2) ON. Observe the LCD and
indicator lights to determine that power is present.

Table of Contents

Other manuals for SciCan Statim 5000

Related product manuals