8. Water Pumps, Reservoir, and Compressor
STATIM 5000/5000S/5000 G4 Service Guide
the top of the line marked MIN in less than that time or if the water level does not reach
the top of the line marked MIN within the allotted time, follow the Pump Recovery Repair
procedures described later in this chapter.
Pump flow parameters
The following pump parameters are valid for this procedure.
Unit Voltage Rating 100 V 50Hz 100 V 60 Hz 110 V 60 Hz 230 V 50 Hz
For units fitted with alluminum steam generators, the following times apply.
Pump type min. max. min. max. min. max. min. max.
Fluid-O-Tech Pump 21.5 23 23 25 N/A N/A 21.5 23
SciCan Pump 18.5 20 21.5 23 21.5 23 21.5 23
Ulka Pump
For units fitted with Stainless Steel steam generators, the following times apply.
SciCan Pump N/A N/A N/A N/A 21.5 23 17.0 18.5
Note: max. / min. numbers represent time in seconds required to fill the bottle to the min. line.