Chapter 1 Installation
GC-2014 Service Manual
6
1.2.2 Installation Surface
Install the instrument on a steady, stable, flat surface.
1.2.3 Corrosive Gas and Dust
In order to attain the specified service life and maintain performance specifications, do
not install the instrument in locations subject to large amounts of corrosive gas or dust.
1.2.4 Influence of Electromagnetic Waves and Power-Supply Noise
Do not use this instrument in locations subject to strong electromagnetic waves or use
with power supplies that generate high noise levels. Doing so may result in malfunction
and it may not be possible to obtain data that conforms to specifications.
1.2.5 Other Precautionary Items
In order to maintain performance specifications, give due consideration to the following
points.
(1) Changes in the ambient temperature during operation must be kept to a minimum.
(2) The instrument must not be directly exposed to the airstream emanating from
heating or cooling equipment.
(3) The instrument must not be exposed to direct sunlight.
(4) Vibrations must be kept to a minimum.