Chapter 3 Gas-Flow Controller
(2) Flow Line for SPL
The flow line shown applies for use in combination with AFC-2014, D-Purge, SPL, and a
manual flow controller.
GC-2014 Service Manual
Resistance tube, air
3-PORT PIPE
Ca
p
illar
y
column
Detecto
r
Column oven
BUFFER ASSY
Tra
p
FILTER ASSY
Tra
p
FILTER ASSY
MOLECULAR SIEVE FILTER
(
Electric S
p
lit Controller
)
Flow-rate senso
r
(
Total Flow Controller
)
Sample
injection unit
(Back of unit)
H
2
(Back of unit)
MAKE UP
(Back of unit)
(
Se
p
tum Pur
g
e Controller
)
Pressure senso
r
Column-inlet
pressure senso
r
(
Back of unit
)
M
H
Resistance tube, H
2
Resistance tube,
make-up gas
Speed
controller
Manual flow controller, H
2
-ai
r
Resistance tube, ai
r
Solenoid-valve assembly
Bellow valve, 2k
Pressure gauge, 200 kPa
Manual flow controller, AUX
73

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2014 Series and is the answer not in the manual?

Shimadzu GC-2014 Series Specifications

General IconGeneral
TypeGas Chromatograph
Detector OptionsFID, TCD, ECD, FPD, FTD
DisplayLCD
Injection PortsUp to 3
Column OptionsCapillary, Packed
Data SystemGCsolution software
WeightApprox. 45 kg
Number of DetectorsUp to 4 detectors
Injection PortSplit/Splitless
Power Requirements100-240V AC, 50/60Hz

Related product manuals