Preparation to use
6-42 MORPHEUS MRI
BY-PASS push button.
Close all three regulators of the flowmeter
box.
Enabled selection control for valves group
enabling (9.4).
Press the BY-PASS (9.2) push button.
Verify that the O
2
flow reaches the valves
group bag.
Enabled fresh gas exit selection control (9.3)
Press the BY-PASS (9.2) push button
Verify that the O
2
flow reaches the TO and
FRO circuit bag connected to the AUX
connector (9.5).
Restore the control enabling the selection of
valves group (9.4).
This function is active also with anaesthesia unit OFF.
Apply the patient circuit to valves group;
INSP.and EXP. connectors
Apply the patient simulator (SIARE type) to
the patient circuit connector.
The suggested patient simulator for tests and checks is the
LS.AB.001 which is equipped with variable compliance and
resistance.