Preparation to use
6-42 MORPHEUS MRI
BY-PASS push button.
• Close all three regulators of the flowmeter
box.
• Enabled selection control for valves group
enabling (9.4).
• Press the BY-PASS (9.2) push button.
• Verify that the O
2
flow reaches the valves
group bag.
• Enabled fresh gas exit selection control (9.3)
• Press the BY-PASS (9.2) push button
• Verify that the O
2
flow reaches the TO and
FRO circuit bag connected to the AUX
connector (9.5).
• Restore the control enabling the selection of
valves group (9.4).
This function is active also with anaesthesia unit OFF.
• Apply the patient circuit to valves group;
INSP.and EXP. connectors
• Apply the patient simulator (SIARE type) to
the patient circuit connector.
The suggested patient simulator for tests and checks is the
LS.AB.001 which is equipped with variable compliance and
resistance.