Preparation to use
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
6-43
• Apen the O
2
(4.11) regulator or AIR (4.6) at
2 l/min.
• Press the AUT key to activate the VAC
ventilation (cfr. 6.5.3)
• The anaesthesia unit begins the inspiratory
(expiratory) phase determining a variation of
parametes measured in the monitoring area
of ventilator.
• Check the regular operation of the
anaethesia unit.
For setting the respiratory physiological parameters (hereinafter
RPP) useful for preliminary tests on lung ventilator, consult the cfr.
6.5.3.
O
2
concentration
• Open the O
2
regulator (4.11) at 7 l/min.
• Verify that the oxygen concentration of
parameters measured in the monitoring area
of ventilator [ O
2
], increases.
• Close the O
2
regulator.
Verify the presence of soda lime in the jar and that it is not
exhausted.