AD.A module
7-10 MORPHEUS MRI
How setting the default respiratory parameters
RATE (Bpm)
30
20
15
I:E
1:2
1:2
1:2
Pmax (cmH2O)
30
30
30
Trigger
-
-
-
VAC
Vt (ml)
0050
0200
0500
PCV
FLOW (l/min)
20
40
60
In STAND-BY modality.
To activate the default settings keep pressed, for a few seconds, the
membrane key relevant to PRESET settings of required respiratory
parameters.
Use of encoder
After having selected the parameter to be
modified, press the key alongside the
parameter.
Turn the encoder clockwise (counter-
clockwise) to increase (decrease) the
parameter value; press the encoder to
confirm the value.