Maintenance
User’s Manual, DU3310103-MRI / rev.3 - 02.05.2019
10-11
Frequency Component Procedure / Action
Several times
a day /
according to
local practice
and standards
Patient circuit Check for any water collection, drain and clean
the tubes when necessary.
Filters
Verify their wear.
Condensation trap
filter
Check for any water collection, drain and clean
when necessary.
Flow sensor Replace according to procedures and local
directives of reference.
Every day /
when
necessary
Oxygen sensor Calibrate according to the procedures
described in this manual.
Flow sensor Replace according to procedures and local
directives of reference.
Condensation trap
filter
Check for any water collection, drain and clean
when necessary.
Anaesthesia unit
General cleaning and checks.
Breathing System Dismount the components and clean, then
sterilize in autoclave, disinfect by steam or
chemically.
Every week /
when
necessary
Breathing System Dismount the components and clean, then
sterilize in autoclave, disinfect by steam or
chemically.
Every 6
months or
1000 working
hours
Anaesthesia unit The anaesthesia unit must be inspected and
checked in general and any worn parts must be
replaced.
Use the appropriate preventive maintenance kit.
This operation must only be carried out by
qualified technical personnel, according to the
instructions contained in the relative service
and maintenance manual.