Maintenance
10-12 MORPHEUS MRI
Frequency Component Procedure / Action
Every 6
months or
1000 working
hours
Oxygen sensor Replace.
The working life of the cell depends on the
working environment. If the temperature or the
O2 % is high, the working life of the sensor will
be lower.
Filters Replace.
Sterilize according to the procedure described
in this manual and according to local standards.
Components that cannot be destroyed should
be sterilised before disposal.
Patient circuit
Washers – O rings
Every year Anaesthesia unit Check the performance.
This includes an electrical safety test and
inspection of the ventilator for mechanical
damage and legibility of the labels.
The anaesthesia unit must also be inspected
and checked in general and worn parts must be
replaced, using the appropriate preventive
maintenance kit.
These operations must only be carried out by
qualified technical personnel, according to the
instructions contained in the relative service
and maintenance manual.
Every two
years / when
necessary
Internal battery Replace.
This operation must only be carried out by
qualified technical personnel, according to the
instructions contained in the relative service
and maintenance manual.
The working life of the battery depends on the
working conditions and environment.
To avoid damage to components due to excessive wear,
carry out preventive maintenance and replace parts
following the recommended frequency.