30 GMS800 · Operating Instructions · 8013025 V1.2 · © SICK MAIHAK GmbH
Start-up
Subject to change without notice
4.1 Safety information on start-up
NOTICE: Risk of damage
Liquids and particles (dust) must not penetrate the gas analyzer measuring
system. The gas analyzer is normally unusable when liquids or particles have
penetrated the measuring system.
Before activating sample gas feed to the gas analyzer:
b
Make sure no liquids (e.g. condensate) or particles can penetrate the gas
analyzer.
b
Check whether sample gas feed to the gas analyzer functions correctly (e.g.
dust filter, valves).
Possible single measures:
b
Wait until system components that remove condensable substances out of
the sample gas
[1]
are in the operating state (e.g. sample gas cooler).
b
Wait until heated system components
[1]
have reached operating temper-
ature (e.g. heated sample gas line).
[1] If fitted