DIGITAL OUTPUT s Point View
A. ZERO CALIBRATION: ZERO CALIBRATION signal output point number
B. SPAN CALIBRATION: SPAN CALIBRATION signal output point number
C. PURGE 1(SV1-SV2): PROBE PURGE1 signal output number
D. PURGE 2(SV3): PROBE PURGE2 signal output number
E. SAMPLING OUTPUT: MEASURING signal output point number
F. REMOT CONTROL OUT: REMOTE CALIBRATION transmission’ signal output
point number
[Picture 21]
TIME PRESET’ s Point View
A. ZERO-CALIB TIME [s]: Time that REMOTE ZERO CALIBRATION is continued;
Unit is [Second],
! Caution: If set value is 0, REMOTE ZERO CALIBRATION does not begin.
B. SPAN-CALIB TIME [s]: Time that REMOTE SPAN CALIBRATION is continued;
Unit is [Second],
! Caution: It set value is 0, REMOTE SPAN CALIBRATION does not begin.
C. CALIB PURBE TIME[s]: After SPAN, normal sampling is retentive during set time.
D. PURGE PERIOD [m]: PURGE cycle time [minute]
E. PURGE DURATION [s]: PURGE time [minute], time of PURGE contact point