74 DUSTHUNTER T · Operating Instructions · 8012428 V 2.0 · © SICK MAIHAK GmbH Germany
Start-up and Parameter Settings
4.4.8 Data backup
All parameters relevant for the collection, processing and input/output of measured values
as well as current measured values can be saved and printed. This allows easy reentering
of set device parameters as needed (e.g. after a firmware update) as well as the
registration of device data or device states for diagnostic purposes.
The following options are available.
● Saving as a project
This allows saving not only device parameters but also data logs.
● Saving as a device file
Stored parameters can be processed without attached devicet and transferred into the
device to a later time again.
● Saving as a protocol
Device data and parameters are recorded in the Parameter protocol.
A Diagnosis protocol can be created for analysis of the device function and recognition
of possible malfunctions.
Saving as a project
At frequent connections to the device we recommend to store a "project". For a renewed
connection it is then only necessary to open this "project". All before stored data are
transmitted automatically into the SOPAS ET.
For saving select the "Project / Export Device" menu and define target directory and file
name. The name of the file to be stored can be chosen freely. It is useful to specify a name
with a reference to the sampling point involved (name of the company, equipment name).
Figure 56 "Project / Export Device" menu
See the Service Manual for a description.