4 START-UP AND PARAMETER SETTINGS
4.5 Setting the Interface module parameters
4.5.1 General information
The following steps are necessary to select and set the optionally available Interface
modules Profibus DP, Modbus TCP and Ethernet type 1:1
Select device fileMCU”, set the measuring system to “Maintenance” mode
Enter the Level 1 password (see “Password and operating levels”, page 86).
Switch to the “Configuration / System Configuration” directory.
The installed interface module is displayed in the field “Interface Module”.
Configure the Interface module according to requirements.
Fig. 56: SOPAS ET menu: MCU/Configuration/System Configuration
GSD file and measured value assignment are available for the Profibus DP module on
request.