108
8012428/YWL2/3-0/2016-08| SICKO P E R A T I N G I N S T R U C T I O N S | DUSTHUNTER T
Subject to change without notice
6 TROUBLESHOOTING
6.3 MCU control unit
6.3.1 Malfunctions
6.3.2 Warning and error messages in the SOPAS ET program
Fig. 76: SOPAS ET menu: MCU/Diagnosis/Error messages/Warnings
● Display of error or warning: With LED symbol
● Description of error or warning: In the description field of SOPAS ET
Symptom Possible cause Action
No display on the LCD
● No supply voltage
● Connection line to LC-Display not
connected or damaged
● Defective fuse
▸ Check voltage supply.
▸ Check the connection line
▸ Exchange fuse.
▸ Contact SICK Customer Service.