Installation
28 FLOWSIC600 · Operating Instructions · 8010458 V2.0 · © SICK MAIHAK GmbH
Bidirectional use
Two straight pipes shall be installed in the inlet and outlet sections if the meter is to be
used bidirectionally. The temperature measuring point shall be disposed downstream the
FLOWSIC600, seen in the direction of predominant use (position 1 or 2 in
→
Figure 7).
Figure 7 FLOWSIC600 installation in the pipeline for bidirectional use
≥ 10 DN FLOWSIC600 ≥ 10 DN
min. 5 DN min. 5 DN
min. 2 DN min. 2 DN min. 2 DN min. 2 DN
1.5 .. 5 DN 1.5 ..5 DN
min. 2 DN min. 7 DN min. 7 DN min. 2 DN
Temperature measuring point
Flow conditioner
Flow conditioner
Temperature measuring point
5 ..8 DN 5 .. 8 DN
DN
Configuration 1
Configuration 2a
Applicable for meters
marked with an asterisk
(*) in
→
»Meter sizes
according to PTB
approval« (pg. 84),
Configuration 2b
Applicable for meters
marked with an asterisk
(*) in
→
»Meter sizes
according to PTB
a p p r o v a l « ( pg . 8 4 ) ,
≥ 5 DN ≥ 5 DN
≥ 10 DN ≥ 10 DN