46 MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG
Operation
Subject to change without notice
Digital inputs
Menu: System Control Unit/Maintenance/Tests/Digital Input
Mark” desired tests and then perform these tests with “Test”.
The following Test menu appears:
Designation Remark
Index Number of the selected input. Shown automatically.
Module Topographic addressing (
“SCU Technical Information“ manual). Shown
automatically.
DI(n)I [State] Computed value of [Source] (“Inverted” is taken into consideration) .
DI(n) [Source] LED off: Physical contact open.
LED on: Physical contact closed.
Index Module Name Inverted
1 N1M01DI01(DI04) di1:Switch Main/Meas
2 N1M01DI02(DI04) di2
3 N1M01DI03(DI04) di3
etc.
Save Mark Test
/System Control Unit/Maintenance/Tests/Digital Input
SCU
MCS100FT
Measure
/System Control Unit/Maintenance/Tests/Digital Input
SCU
MCS100FT
Measure
1Index
DI(n) [Source]
N1M01DI01(DI04)Module
DI(n)I [State]
Cancel

Table of Contents