Operation
MCS100FT · Operating Instructions · 8011893 V2-1 · © SICK AG 47
Subject to change without notice
Digital outputs
Menu: System Control Unit/Maintenance/Tests/Digital Output.
Mark” desired tests and then perform these tests with “Test”.
The following Test menu appears:
Designation Remark
Index Number of the selected output. Shown automatically.
Module Topographic addressing (
“SCU Technical Information“ manual). Shown
automatically.
Test Parameter No checkmark: Physical contact should be open.
Checkmark: Physical contact should be closed.
DO(n)O [State] LED off: Relay open.
LED on: Relay closed.
DO(n) [Source] LED off: Program specification: Physical contact should be open.
LED on: Program specification: Physical contact should be closed.
/System Control Unit/Maintenance/Tests/Digital Output
SCU
MCS100FT
Measure
Index Module Source Inverted
1 N1M02DO01(DO04) bv11
2 N1M02DO02(DO04) bv12
3 N1M02DO03(DO04) s2e9
etc.
Save Mark Test
/System Control Unit/Maintenance/Tests/Digital Output
SCU
MCS100FT
Measure
1Index
DO(n) [Source]
N1M02DO01(DO04)Module
DO(n)O [State]
Cancel
Test Parameter