2-22
1010PFM-3GSection 2
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index (631) 248 77774
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? Yes
Empty Pipe Set Channel Not Setup
Zero Flow Adjust Channel Not Setup
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Press <ENT>. The [Install Completed?] menu cell will indicate [Yes] after the 1010 is successfully
installed.
Force Transmit Procedure
This diagnostic software routine allows the user to “force” a transmitting condition that can be use to
search for an amplitude level (ALC) when Detection Fault or Low Signal alarms are present. The
routine forces the flowmeter to generate constant transmit bursts while reporting current receive signal
strength for the user. To initiate the Force Transmit function, refer to the example shown below.
1. After [Install] is selected press <ENT>. While the meter is going through the drives (see menu
screen below), press the <ALT> and <MENU> keys simultaneously.
To complete the Install process, disregard the following paragraphs explaining the Force Trans-
mit and Force Frequency diagnostic software routines.
NOTE: If the transducers have been installed successfully but the Estimated Vs (sonic veloc-
ity) has been changed, the Pick/Install Xdcr menu cell [Install Completed?] option list
will also display the [Transfer Install] function selection. The Transfer Install function
allows the transducers to be optimally positioned for a different fluid, without the need
for a new Initial Makeup procedure (see the Water Calibration addendum in this manual
for details).
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? No
Empty Pipe Set Channel Not Setup
Zero Flow Adjust Channel Not Setup
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? Install
Empty Pipe Set Channel Not Setup
Zero Flow Adjust Channel Not Setup
Siemens 2 Channel [1] SITE1
Drive 11
Pick/Install Xdcr