2-76
1010PFM-3GSection 2
¦ Access the [Pick/Install Xdcr] menu. Check the [Transducer Size] menu cell. If necessary, enter
the [Xdcr Size] option list and pick the transducer under test.
¦ Check the [Xdcr Mount Mode] menu cell. Adjust to match the simulator chart above.
¦ Press the <Down Arrow> to move the highlight down to [Install Completed?]. Access the option list.
Press the Up or Down Arrows> and move the cursor to [Install]. Press <ENT> to start the trans-
ducer install routine.
NOTE: Since Sizes A and C transducers are installed in Reflect Mode, you will see a pop-up
window that prompts you to: [Use Actual Zero]. You can ignore this by pressing the
<Down Arrow>.
Upon the completion of the [Install] procedure, the computer should report a sonic velocity within the
range of approximately 1350m/s to 1700 m/s (depending on the transducer size under test). Next,
check the [Diagnostic Data] menu. The [Valc %] item must be >35 for dependable operation. In
addition, note the reading, then compare it to the [Vf max] item in the [Diagnostic Data/Site Setup
Data] menu. The value of the reading should be less that 2% of the published Vf max. Confirming
these values certifies that the entire system (computer, transducers, cables) is operating correctly.
The investigation should proceed to a review of all site conditions to locate the operating problem.
2.11.9 USING THE 996PSP PIPE SIMULATOR
Using the proper coupling compound, mount the transducers on the pipe simulator as shown on the
next page. Slide each transducer until it presses against a pin-stop. Use the clamping screws to hold
the transducers in place.
¦ Access the Installation Menu. Select [Meter Type] [Single, Dual, or Quad Channel] depending on
meter type. Select the meter channel (1,2,3, or 4) depending on which measurement channel you
intend to test. Select [Clamp-On] and then [Channel Setup].
¦ Access the [Channel/Path Setup] menu. Move the highlight to [Create/Name Site]. Create a new
Site Setup (e.g., TEST1). You can now enter data without altering an existing Site Setup.
¦ Select the [Pipe Data] Menu. Referencing either the English or metric pipe simulator chart below,
enter the pipe data corresponding to the transducer size under test.
Part Xdcr Pipe OD Pipe Wall Mount Spacing Number
Number Size (in) Mat’l Thk (in) Mode Offset Index
1012TB-1 A 0.650 Steel* 0.100* Reflect Nominal 7
1012TB-1 B 1.150 Steel* 0.100* Direct Minimum 4
1012TB-2 C 2.000 Steel* 0.100* Reflect Nominal 11
1012TB-2 D 3.500 Steel* 0.100* Direct Minimum Use Ltn
1012TB-1 & 2 Universal Transducer Test Block Chart (English)
*System Defaults
1012TB-1 A 16.5 Steel* 2.54* Reflect Nominal 7
1012TB-1 B 29.2 Steel* 2.54* Direct Minimum 4
1012TB-2 C 50.8 Steel* 2.54* Reflect Nominal 11
1012TB-2 D 88.9 Steel* 2.54* Direct Minimum Use Ltn
* System Defaults
1012TB-1 and 2 Universal Transducer Test Block Chart (Metric)
Part Xdcr Pipe OD Pipe Wall Mount Spacing Number
Number Size (mm) Mat’l Thk (mm) Mode Offset Index