2-17
1010PFM-3G
Section 2
2.4.1 HOW TO SELECT A TRANSDUCER MODEL
Use the [Transducer Model] menu cell to define the type of transducer for use with your application.
This allows the flow computer to adjust its transmit/receive functions accordingly. You can choose
from either the [1011 Universal] or [1011H High Precision] transducer lists, or from a list of [991 Univer-
sal] models. You should consider using high precision transducers for extremely critical applications.
We usually recommend these transducers for custody transfer, leak detection or nuclear power appli-
cations. The system default is [1011] Universal]. If this suits the application, bypass this menu cell by
pressing .
PICK/INSTALL XDCR MENU STRUCTURE
Pick/Install Xdcr Transducer Model 991 Universal
1011 Universal
1011H Hi Precision
Transducer Size
A1A1 0A0A A1HA1H
A2 0 A2H
B1 1 A3H
B2 2A B1H
B3 2 B2H
C1 3A C1H
C2 3 C2H
C3 4A D1H
D1 4 D2H
D2 5A D4H
D3 5
E1
E2
E3
Xdcr Mount Mode Direct
Reflect
Spacing Offset Minimum
Nominal
Maximum
Number Index 4 (generated)
Spacing Method Spacer Bar [P/N] *auto P/N generation
Ltn Value 2.00 (generated)
Install Completed? No [Yes] *[Yes] indicates successful install
Install
Empty Pipe Set MTYmatic
Set Empty
Actual MTY
Zero Flow Adjust AutoZero *Reflect mount only
ZeroClr
Reversamatic
Actual Zero
Zeromatic
1011 Universal 991 Universal 1011H High Precision