2-21
1010PFM-3G
Section 2
2.4.9 HOW TO USE [INSTALL COMPLETED?]
Use the [Install Completed?] menu cell to inform the flow computer that you have mounted the trans-
ducers according to the selected mode and spacing requirements and are ready to start operation.
To start the [Install Completed?] routine for an original installation:
Select any transit-time mode (e.g., Clamp-On or Flow Tube) and press the <Right Arrow>.
Press <Down Arrow> and scroll to the [Pick/Install Xdcr] menu cell. Press <Right Arrow>.
Install transducers as required (refer to Section 3 for procedures).
After transducers are properly mounted, the flow computer completes its Initial Makeup command and
the [Install Completed?] menu cell appears.
To access the [Install Completed?] option list, press the <Right Arrow>.
Press <Down Arrow> and scroll to the [Install Completed] menu cell and then press <ENT> (un-
less otherwise directed to do so by the Technical Service Department).
Transducer Model 1011 Universal
Transducer Size B3
Xdcr Mount Mode Direct
Spacing Offset Minimum
Number Index 4
Spacing Method Track 1012TP
Ltn Value (in) 0.217
Install Completed? No
Empty Pipe Set Channel Not Setup
Zero Flow Adjust Channel Not Setup
Siemens 2 Channel [1] SITE1
Key [Install] after mounting transducers
Pick/Install Xdcr
NOTES
z Mount the transducers using the selected mode (direct or reflect). Please refer to the trans-
ducer mounting procedures detailed in Section 3.
z When using a mounting track or a spacer bar in reflect mode, locate the first transducer at the
Reference Index and the second transducer at the recommended Number Index.
z If you are mounting the transducers independently, you must use the recommended distance;
Ltn to space the transducers.
z You must use the proper sonic coupling compound. See “Recommended Sonic Coupling
Compounds” in Section 5 for appropriate type and part number.
z The pipe must be completely filled with a liquid, which can be either flowing or at zero flow.