1-19
1010PFM-3G
Section 1
Press . Note that the cursor changes to a block, which appears after a question mark (?). This
indicates that you have accessed the first character position of an eight-character (maximum) alpha-
numeric entry field.
Creating a New Site Setup
(See also paragraph 2.1.3)
The Channel Setup section of this manual provides detailed instructions on how to use all the Channel
Setup menu commands. This section provides an example of how to use the [Create/Rename Site]
command to create a new site setup. Always begin a new installation by issuing the [Create/Name
Site] command. Although you do not need to enter a site name to create a new site setup, you have to
provide one to save the site setup or to identify the source of logged data. In this example, we will use
[SITE1] for the site name.
To access the [Channel Setup] menu press then move the highlight down to [Create/Name Site]
with the key.
To select [Clamp-on] press . This selects the Clamp-on installation menu for Measurement Chan-
nel 1. Note that the first menu, [Channel Setup], is highlighted.
Channel 1 >Clamp-on
Channel 2 FlowTube
Reflexor
Siemens 2 Channel [1] Channel 1
Choose Channel 1 Flowmeter Type
Channel Setup
Siemens 2 Channel [1] Channel 1
Recall Site Setup No Sites
Channel Enable No
Create/Name Site
Site Security Off
Delete Site Setup No Sites
Save/Rename Site
Right Arrow & Enter Creates a new Site
Channel Setup