9.2.6.2 Adjusting the zero point (differential pressure)
Requirement
The pressure measurement is stable.
Procedure
1. Make sure there is identical pressure in the two process connections.
2.
Navigate to the parameter view.
Navigating in the views (Page 76)
3. In the parameter view, select the "Zero point adjustment" parameter [07].
4. Press the
button.
5. Set the zero point to 0 or to the desired value.
6.
Confirm the value by pressing the
button.
7. Change to the measurement view with the button.
Note
Depending on the damping setting, a settling time elapses until the pressure measurement 0
is displayed.
Make sure there is identical pressure in the two process connections until the end of the
operation.
Result
The device displays pressure measurement 0 in the set unit.
The effective measuring range is reduced by the amount of the upstream pressure.
Example: At a pre-load pressure of 25 mbar, the upper measuring range limit of a 250 mbar
pressure transmitter is reduced to 225 mbar.
9.2.6.3 Adjusting zero point (absolute pressure)
Requirement
You have created a reference pressure that is within the measurement limits.
Procedure
1. Navigate to the parameter view.
Navigating in the views (Page 76)
2. Select the "Zero point adjustment" parameter [07].
Parameter assignment
9.2 Parameter assignment over device with display
SITRANS P320/P420 with 4 to 20 mA/HART
Operating Instructions, 11/2019, A5E44852162-AC 121