13.6 Vibration resistance
General operating conditions Gauge pressure series
2)
Aluminum and stainless steel enclosure
Differential pressure series
1)
Aluminum and stainless steel enclosure
Vibrations (sine)
IEC 60068-2-6
2 ... 9 Hz at 0.3 mm
9 ... 200 Hz at 5 m/s²
1 octave/min
5 cycles/axis
Continuous shocks (half-sine)
IEC 60068-2-27
70 m/s²
30 ms
6 shocks/axis
Continuous shocks (half-sine)
IEC 60068-2-27
250 m/s²
6 ms
1000 shocks/axis
1) Without mounting bracket
2) With mounting bracket
Operating conditions according to KTA
3503
Gauge pressure series
2)
Aluminum and stainless steel enclosure
Differential pressure series
2)
Aluminum and stainless steel enclosure
Vibrations (sine)
IEC 60068-2-6
9 ... 35 Hz at 10 m/s²
1 octave/min
1 cycle/axis
Vibrations (sine)
IEC 60068-2-6
5 ... 7 Hz at 20 mm
9 ... 100 Hz at 20 m/s²
10 octave/min
1 cycle/axis
Continuous shocks (half-sine)
IEC 60068-2-27
300 m/s²
11 ms
6 shocks/axis
2) With mounting bracket
Operating conditions according to IEC
61298-3 (2g normal)
Gauge pressure series
2)
Aluminum and stainless steel enclosure
Differential pressure series
1)2)
Aluminum and stainless steel enclosure
Vibrations (sine)
IEC 60068-2-6
10 ... 58 Hz at 0.3 mm
58 ... 1000 Hz at 20 m/s²
1 octave/min
20 cycles/axis
1) Without mounting bracket
2) With mounting bracket
Technical data
13.6 Vibration resistance
SITRANS P320/P420 with 4 to 20 mA/HART
Operating Instructions, 11/2019, A5E44852162-AC 233