Safety Software Related Safety Information
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
A.4-31
0.0
A.4
A.4
A.4
CAUTION
A.4
User enters wrong values for the calibration factors!
A.4
Wrong Calcium Scoring results may lead to wrong diagnosis.
A. 4
✧ Only use correct calibration factor values, check the input
carefully.
A.4
CAUTION
A.4
The scanner may not be correctly identified by Calcium Scoring
during image load when using "remembered" calibration fac-
tors!
A.4
Wrong calibration factors may lead to wrong results.
A. 4
✧ Carefully check the suggested factors and make sure that
they belong to the used scanner, before accepting the val-
ues.