Use of Accessories
Using the table in/out joystick
C.4
You can move the tabletop horizontally with the joystick
mounted at the patient table.
C.4
Mounting the joystick
C.4
✓ On each side of the patient table an accessory rail is pro-
vided for attaching the clamping device for taking the joy-
stick.
✧ Hold the lever of the clamping device pressed when attach-
ing it to the accessory rail.
✧ Lock the joystick in the two inserts of the clamping device
holding the push buttons pressed.
✧ Connect the joystick plug to the corresponding socket.
✧ Check the function of the joystick.
The direction of table and joystick movement must correspond
with each other.
C.4
B
.4
C
A schematic representation of the gantry and the table on both,
the clamping device and the joystick, serves as an orientation
aid for correct positioning. Ensure that the gantry symbol points
towards the gantry when mounting both accessory parts.