EasyManua.ls Logo

Siemens SOMATOM Emotion syngo CT 2007E - Shutdown in Case of Emergency

Siemens SOMATOM Emotion syngo CT 2007E
574 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
Information about Personal Safety Safety
A.2-20
Shutdown in case of emergency
A.2
In addition to the STOP keys at the CT unit, EMERGENCY OFF but-
tons to be provided by the customer have to be easily visible and
within easy reach. Country specific regulations must be
observed.
A.2
B
.2
A
In extreme cases, you can press an EMERGENCY OFF button to
shut down the whole system. Radiation and system movements
will be stopped.
A.2
If the power supply of the SOMATOM Emotion computer com-
ponents is not integrated in the EMERGENCY OFF system, the
UPS must be unplugged to disconnect the computer compo-
nents from main power.
A.2
You must only press the EMERGENCY OFF button:
A.2
If the STOP button on the CT system does not respond.
If situations arise that could cause injury to the patient or
user or damage to the equipment (e.g., liquids or particles
getting into the equipment, damage of cover parts, fire).

Table of Contents