syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.5-1
0.0
CHAPTER
F.4
F.5
Low Contrast Test
(Optional)
This chapter describes how to use the low contrast section of
the phantom set and how to do the low contrast test.
F.5
B
.5
F
The availability of the low contrast test is country-specific. It has
to be activated by Siemens Service.
F.5
Low contrast phantom section
F.5
The low contrast phantom section consists of a 165 mm diam-
eter, 25 mm thick, water-equivalent, brown plastic cylinder. A
clear plastic insert at the center contains 3 sets of low contrast
pins (with diameters of 3, 4, 5 mm, each set with 4 pins),
together with two 20 mm diameter measurement areas.
F.5
Schematic diagram
F.5
The two 20 mm diameter measurement areas are used to estab-
lish low contrast differences between the pins and the sur-
rounding area. The rated contrast between the pins and the sur-
rounding area is 6 HU 1 HU (0.6 % 0.1 % contrast).
F.5
F.5