Quality Assurance Camera Test
0.0
syngo CT 2007E Emotion Operator Manual
02/2008
0.0
F.6-5
0.0
Evaluating the test
F.6
✧ Compare the test results with the reference values.
✧ Measure the fields marked in the first test image with the
densitometer X-RITE 331.
✧ Enter the measured values in the constancy test report.
fB
.6
F
Test is out of tolerance.
F.6
✧ Contact Siemens Customer Service.
F.6
Example
F.6
Here is an example of test image 1 with measurement fields
marked by the technologist:
F.6
F.6